Product Description:
Didymium glass has been used for many years as a high wavelength visible reference material. When prepared in perchloric acid, didymium gives a spectral scan containing a series of sharp characteristic peaks that extend well above the useable range of the holmium into the near NIR (680-900 nm). Again, like holmium, these values are dependent and will vary with the spectral bandwidth of the measuring instrument.
|
 |
Didymium - UV and Visible Wavelength
Description: Didymium (neodymium
& praesodymium) in perchloric acid.
Primary Usage: Assessment of wavelength scale accuracy in both UV and visible regions.
Useable range: 290 nm to 870 nm, instruments with SBW of less than 5 nm.
Physical Configuration: Far UV quartz cells that have been permanently heat sealed. |
 |
 |
 |
| 0.10 |
864.35 |
799.18 |
794.14 |
740.06 |
731.57 |
575.19 |
521.82 |
511.81 |
481.65 |
468.78 |
444.05 |
353.83 |
328.92 |
298.24 |
| 0.25 |
864.32 |
799.63 |
794.16 |
740.08 |
731.57 |
575.14 |
521.81 |
511.83 |
481.64 |
468.80 |
444.06 |
353.82 |
328.99 |
298.25 |
| 0.50 |
864.32 |
799.63 |
794.12 |
740.02 |
731.56 |
575.12 |
521.76 |
511.85 |
481.63 |
468.76 |
444.04 |
353.81 |
328.74 |
298.24 |
| 1.00 |
864.40 |
798.99 |
794.08 |
740.04 |
731.63 |
575.04 |
521.56 |
511.87 |
481.67 |
468.66 |
444.01 |
353.80 |
328.78 |
298.29 |
| 1.50 |
864.34 |
799.83 |
794.09 |
740.02 |
731.74 |
574.87 |
521.38 |
511.90 |
481.70 |
468.61 |
443.96 |
353.82 |
328.79 |
298.51 |
| 2.00 |
864.37 |
799.63 |
794.09 |
740.17 |
731.92 |
574.87 |
521.38 |
511.90 |
481.70 |
468.48 |
443.88 |
353.83 |
328.82 |
298.52 |
| 3.00 |
864.37 |
799.63 |
794.09 |
740.34 |
732.48 |
574.98 |
521.38 |
511.57 |
481.64 |
468.25 |
443.73 |
353.53 |
328.84 |
298.79 |
|
 |
 |
 |
 |
 |
All Starna® didymium Certified Reference Materials are manufactured using procedures similar to those used in the preparation of our Holmium Oxide Reference Material (RM-HL). After filling under controlled conditions, the cells are then permanently sealed by heat fusion and the values certified by the procedure described below.
Calibration procedure
Traceability:
. Primary instrumental wavelength calibration is established using the emission lines from mercury and deuterium sources.
.Additional traceability links to NIST primary materials are established using SRM 2034 holmium oxide (4% m/v) in perchloric acid (10% v/v). |
| |
 |
Use:
All appropriate fundamental parameters and procedures relating to measurement, handling and storage are fully documented on the certificate supplied with each Certified Reference Material. |
|
| |
 |
|
 |
 |
 |