Description: Samarium (III) oxide in perchloric acid.
Primary Usage: Assessment of wavelength scale accuracy in both UV and visible regions.
Useable range: 230 nm to 560 nm, instruments with SBW of less than 5 nm.
Physical Configuration: Far UV quartz cells that have been permanently heat sealed.
RM-SL: (see MSDS)
Consists of one sealed cell, with certified peak at spectral bandwidth values of 0.10 nm, 0.25 nm, 0.50 nm, 1.00 nm, 1.50 nm, 2.00 nm and 3.00 nm. |
 |
Product Description:
Samarium perchlorate is a particularly suitable reference material for checking the wavelength scale of a spectrophotometer over the most commonly used range of 200 to 500 nm, as it has peaks throughout this region. Many of the peaks are very narrow, providing very accurate location of the peak wavelengths. With a half bandwidth of as little as 5 nm, some peaks provide a very convenient way of checking instrument resolution also. Shown below is this effect observed on the peak couple at 233-235 nm. |