Purpose
These references can be used to qualify the absorbance accuracy and linearity of near-infrared spectrophotometers with a spectral bandwidth of 6.5 nm or less, over the wavelength range 800 nm - 3200 nm.
Description and Discussion
Glass filters mounted ‘stress free’ in an anodised aluminiumholder. The use of Schott NG-type glasses is well established for the validation of the absorbance scale and linearity of spectrophotometers in the near-infrared region.
The spectrum of these filters is relatively flat over most of the range of calibration, but with significant deviation above 2700 n
Calibration values (in % Transmission and Absorbance) are certified at 1100, 1700, 2210, 2500 and 2850 nm. Three filters are available, designated, R1 R3 and I1. Approximate values are:
| Wavelength | 1100 nm | 1700 nm | 2210 nm | 2500 nm | 2850 nm | |||||
| Scale | %T | A | %T | A | %T | A | %T | A | %T | A |
| R1 | 20 | 0.70 | 40 | 0.40 | 47 | 0.32 | 49 | 0.31 | 2 | 1.83 |
| R3 | 25 | 0.59 | 46 | 0.34 | 53 | 0.28 | 55 | 0.26 | 3 | 1.54 |
| I1 | 33 | 0.48 | 53 | 0.28 | 59 | 0.23 | 60 | 0.22 | 6 | 1.25 |